S3 Cell Sorter Instruction Manual
| 79
Shutdown
8
8.1 Daily Shutdown
The S3 system should be shut down at the end of the day. This will turn off all droplet
formation and charging, lasers, and streams. A flush will be completed with the DI water
container to rinse the system.
IMPORTANT!
Always follow the personal protective equipment (PPE) guidelines
relevant to your laboratory’s safety procedures for dealing with ethanol or bleach.
Note:
At any time, the shutdown procedure can be canceled by clicking the X icon on the
dialog windows.
Note:
If a user chooses to log off from the software, the force cleaning procedure option can
be run. Similar to the shutdown procedure, windows will appear with the exception of the
automatic startup window.
To shut down:
1. Ensure that the sample loading stage is in the wash position.
2. On the Setup and Maintenance tab select
Shutdown
. The software will confirm with a
dialog box that you intend to shut down the system (Figure 88).
Fig. 88. System shutdown check.
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......