7 Main Menu Device Setup
LB 470RID Level
112
56925-1BA2 Rev.02, 05/2021
9.
Enter the absorption coefficient of the product. Enter it with a negative pre-
fix. Unless it is one of the very rare applications with an inverted
characteristic.
10.
Enter the product density.
11.
Enter the measurement path.
12.
Click on the button
<Calculate>
.
The second Point was calculated.
13.
Click on
<OK>
to confirm the new calibration point.
Calibration Settings: Chart
Device Setup | Setup | Calibration | Calibration Settings
The characteristic curve of the calibration performed is shown in the tab “Chart”.
1
2
3
4
Axis level [%}
Axis count rate [cps]
Characteristic chart (exponential)
Characteristic chart (linear)
Fig. 83
Calibration Parameters (Chart)