LB 470RID Level
7 Main Menu Device Setup
56925-1BA2 Rev.02, 05/2021
113
Calibrate
Device Setup | Setup | Calibration
Data that are necessary for a complete measurement are found in the calibration
parameter set. All the data of the calibration parameter set are transferred to the
measurement parameter set when the button “Calibrate” is clicked. Only after this
are they can be used for measurement value calculation.
1
2
3
Button
<Calibrate>
Button
<Yes>
Button
<Close>
Fig. 84
Calibrate
1.
Click on the button
<Calibrate>
(Fig. 84, item 1).
A new window with the message “Use calibration set for measurement” ap-
pears.
2.
Click on the
<Yes>
button (Fig. 84, item 2) to overwrite your measurement
parameters with the calibration settings.
A new window with the message “Calibration successful” appears.
3.
Click on the
<Close>
button (Fig. 84, item 3)
The calibration was performed.
4.
Activate the check box in the
Device Settings | Setup | Signal processing | RID
.
NOTICE
Check your calibration by simulating a detector count rate. Use the test count rate
in the simulation menu.
As the value of the test count rates, e.g. the counting rates from the calibration
points are used. Check whether the correct measured value is displayed at the
respective test count rate.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...