Calibration Settings: Background
Device Setup | Setup | Calibration | Calibration Settings
The background count rate (Fig. 76, item 1) is the natural background radiation
seen by the detector and must be measured at least for rod detectors. This count
rate is compensated for by the system. The best approach for your background
measurement is dependent on the situation on site and on the type of radiation
source.
NOTICE
A closed shield also results in measurable residual radiation, which can falsify the
measurement of background radiation. For this reason, it is recommended that
the detector (Fig. 75, item 4) be determined at a suitable distance (approx. 10 m)
or behind a thick concrete wall (Fig. 75, item 2).
In the case of measuring arrangements with point sources, it is recommended
to place the shield with spotlights at a suitable distance (approx. 10 m), or
behind a thick concrete wall.
To avoid calibration errors, it must be ensured that during calibration and
commissioning of the measuring equipment no welding tests are being done
on the site, even at long distance.
1
2
3
4
Radiation shield (closed)
Concrete wall
Distance
Rod detector
Fig. 75
Conditions during background determination
IMPORTANT
All entries and changes in the “Parameters” tab will take effect only when you
click on the
<Calibrate>
button in the submenu “Calibrate”.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...