7 Main Menu Device Setup
LB 470RID Level
68
56925-1BA2 Rev.02, 05/2021
Perform MU Update
1.
Save the current update file of the MU software on a USB storage device.
2.
Connect a USB storage device to the front of device (Fig. 4, item 5).
3.
In the “Device Setup” menu, click on “Device Information” (
Device Setup |
Identification | Device Information
).
4.
The USB storage device is recognized by the system after a few seconds and
the
<MU Update>
(Fig. 35, item 2) button can be clicked.
The selection window “update file” (Fig. 35, item 6) opens.
5.
Select the appropriate file and click on the button
<MU Update>
item 2). Confirm with
<Yes>
.
The update is performed and the measurement is interrupted.
6.
Click the Button
<Restart>
to reboot the EVU.
The device restarts and the new MU software has been installed.
NOTICE
Berthold recommends calibrating the current outputs whenever a module has
been installed/replaced or if a software update has been carried out.