7 Main Menu Device Setup
LB 470RID Level
108
56925-1BA2 Rev.02, 05/2021
Calibration Settings: PV Range
Device Setup | Setup | Calibration | Calibration Settings
1
2
Input field “PV lower range value” [%]
Input field “PV upper range value” [%]
Fig. 79
Calibration Parameters (PV Range)
PV Range Value
Lower limit of the measuring range. Is also used as lower
limit of the current output during calibration. These can
subsequently be adjusted separately (see chapter 7.3.5
Signal Condition).
PV Range Value
Upper limit of the measuring range. Is also used as lower
limit of the current output during calibration. These can
subsequently be adjusted separately (see chapter 7.3.5
Signal Condition).
During calibration, the count rates of both channels (measurement channel and
secondary channel) are recorded.