9 Troubleshooting
LB 470RID Level
178
56925-1BA2 Rev.02, 05/2021
9.2.1
System
Code
Message
NAMUR107 Help Text
M101
HW Module
F
Hardware electronics module corrupt.
Restart the device. Contact Berthold ser-
vice, if this event occurs repeatedly.
M102
Device dataset
F
Failure of the permanent memory. No
parameter set found. Factory reset and /
or restart the device. Contact Berthold
service, if this event occurs repeatedly.
The device possibly must be reset twice.
M103
RAM, Flash or CPU
F
Internal hardware failure. Restart the
device. Contact Berthold service, if this
event occurs repeatedly.
M104
WD Reboot
M
The Watchdog has caused the device to
restart. Contact Berthold service, if this
event occurs repeatedly. Check, if mas-
sive electromagnetic interferences have
caused this event.
M105
WD Failure
F
Watchdog malfunction. Contact
Berthold service, if this event occurs re-
peatedly.
M106
WD Off
M
Watchdog is inactive. Activate Watch-
dog
M107
Error in the internal real
time clock
M
Malfunction of the real-time clock.
Check Date and Time. If the event oc-
curs frequently, contact Berthold
Service.
M108
CPU temperature sensor
M
The temperature sensor of the device is
defective. Contact Berthold Service. The
hardware is defective and, if necessary,
must be checked and replaced.
M109
Lower temperature limit:
Maintenance required
M
The internal temperature of the device
is close to the lower threshold value
(-20°C) of the permissible operating
temperature.
M110
Temp LL OOS
S
The internal temperature of the device
is below the lower limit (-30°C). The cor-
rect function of the device cannot be
guaranteed. It is recommend to have
the device checked by Berthold Service,
even if it seems to work normally.
M111
Temp UL maintenance
M
The internal temperature of the device
is close to the upper limit (70°C).
M112
Temp UL OOS
S
The internal temperature of the device
is above the upper limit (85°C). The cor-
rect function of the device cannot be
guaranteed. It is recommend to have
the device checked by Berthold Service,
even if it seems to work normally.
M113
Power On Reboot
C
The device was restarted, e.g. due to a
power failure.
M114
Software Reboot
C
The device was restarted by user input.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...