LB 470RID Level
5 Electric Installation
56925-1BA2 Rev.02, 05/2021
45
only use cables whose diameters are approved for the respective cable gland.
The cables must comply with the requirements and cross-sections specified in
the technical data.
The connected cables must be suitable for a temperature that is at least 10°C
above the maximum permissible ambient temperature.
5.1.3
Cable Glands and Blanking Elements
The feeding of cables into the wall housing is only permitted via a cable entry.
Cable glands must be suitable for the respective application.
All cable glands must be assembled according to manufacturer's instructions
and be tightened to the appropriate tightening torque.
Cable glands that are not required for installation must be covered with suit-
able blanking elements.
Line cross-sections must comply with the respectively used cables.
Cable bushings and blanking elements must comply with the applicable IP pro-
tection class and with the requirements for the operational environment.
We recommend ordering missing cable glands, blanking elements or adapters
from Berthold.