The recorded values are read and entered into the table (Fig. 62), the plateau
curve (Fig. 63) is drawn and stored automatically.
Plateau Table
Device Setup | Setup | Sensors | [NAME DETECTOR] | Plateau | Plateau Table
The data from each measurement point are listed in the plateau table. The data
from the plateau table can be exported to a USB memory device.
1
2
3
4
5
Date and time of measurement recording
High voltage (HV) in volts
Count rate in the measurement channel (cps)
Button
<Export Plateau Data>
Window “Export successful”
Fig. 62
Plateau Table
Export Plateau Data
1.
Connect a USB flash drive to the USB port of the EVU.
The USB memory device is recognized by the system after a few seconds and
the button
<Export Plateau Data>
can be clicked.
2.
Click on the button
<Export Plateau Data>
(Fig. 62, item 4).
The values of the plateau measurement have been stored in a .txt file.
3.
Confirm the message with
<OK>
.
Information
The file name is derived from "Plateau", the date and time of the measurement
process (PlateauYYYYMMDD_hr_min_sec.txt).
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...