7 Main Menu Device Setup
LB 470RID Level
128
56925-1BA2 Rev.02, 05/2021
GPC Nuclide Settings
Device Setup | Setup | Calibration | GPC | GPC Nuclide Settings
Under "GPC Settings" the isotope of the source for gas density measurement can
be selected (Fig. 98, item 1). The half-life time of the isotope is shown in the display
field (Fig. 98, item 2).
NOTICE
The sources for level measurement and the sources of the GPC measurement
must contain the same Nuclide (see
Device Setup | Setup | Calibration | Calibra-
tion Settings
tab “Nuclide).
1
2
Selection arrow “Nuclide”
Display field “Half-life” [years]
Fig. 98
GPC Nuclide Settings
Reset GPC Settings
Device Setup | Setup | Calibration | GPC | Reset GPC
All GPC settings can be reset in the window “Reset GPC Settings
″
.
1
2
Button
<Reset GPC settings>
Window “GPC Reset Warning”
Fig. 99
Reset GPC Settings
1.
Click on the button
<Reset GPC settings>
(Fig. 99, item 1).
A confirmation message (Fig. 99, item 2) appears.
2.
Click on
<Yes>
to set all values to “Default”, click on
<No>
to cancel.