7 Main Menu Device Setup
LB 470RID Level
110
56925-1BA2 Rev.02, 05/2021
Calibration Settings: Table (2 Point Exp. calibration type)
Device Setup | Setup | Calibration | Calibration Settings
1
2
3
4
5
6
7
Column count rate [%]
Column count rate [cps]
Column count rate [rid]
Button
<Edit>
Button Add measuring point
< + >
Button Delete point
< - >
Date of the last editing
8
9
10
11
12
13
Count of the calibration points
Read-In-Time [s]
Count rate RID [cps]
Count rate [cps]
Process value [%]
Measuring process (determine
new calibration point)
Fig. 81
Calibration parameters (Table 2 Point Exp. calibration type)
NOTICE
To avoid calibration errors, it must be ensured that during calibration and com-
missioning of the measuring equipment no welding tests are being done on the
site, even at long distance.
Add new Calibration Point
1.
Click on the button
< + >
(Fig. 81, item 5).
The window “New calibration point” (Fig. 81, item 9-12) opens.
2.
Click on the input field “Process value” (Fig. 81, item 12) to open the input
field.
3.
Specify the current, actual level in percent and confirm with the Enter key.
4.
Click on the input field “Read-In Time” (Fig. 81, item 9) to open the input
field.
5.
Specify a read-in time and confirm with the Enter key.
6.
Click on the button
<Start>.
The Window “New calibration Point” (Fig. 81, item 13) opens and the count
rate is determined.
7.
Click on
<OK>
to confirm the new calibration point.
The new calibration point is recorded in the table.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...