![Berthold DUO XPERT LB 470RID Operating Manual Download Page 73](http://html1.mh-extra.com/html/berthold/duo-xpert-lb-470rid/duo-xpert-lb-470rid_operating-manual_2739258073.webp)
LB 470RID Level
7 Main Menu Device Setup
56925-1BA2 Rev.02, 05/2021
73
Set Date and Time
Device Setup | Setup | System | Date / Time
IMPORTANT
The date and time must always be set correctly so that all records (log files) have
the correct metadata. The correct date is also indispensable for the decay com-
pensation.
1
2
3
4
Display date
Display time
Button
<Apply>
Arrow key Time
5
6
7
8
Selection arrow Date
Calendar
Arrow keys Month
Input box Year
Fig. 40
Date / Time, calendar
1.
Click on the arrow key (Fig. 40, item 5) in order to set the date.
The calendar is opened (Fig. 40, item 6).
2.
Click on the year number (Fig. 40, item 8) in order to enter the year.
3.
Set the month (Fig. 40, item 7) by clicking on the arrow keys.
4.
Set the day by clicking on a number in the calendar.
5.
Change the time by clicking on the arrow keys (Fig. 40, item 4).
6.
Click on
<Apply>
(Fig. 40, item 3), to accept the date and time settings.
Date and the time have been set
NOTICE
The real-time clock for date and time is buffered via a capacitor and continues to
run for up to approx. 4 weeks even when the device is switched off.
If the device has been out of operation for more than 4 weeks, error M116
appears. The date and time must then be reset.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...