7 Main Menu Device Setup
LB 470RID Level
116
56925-1BA2 Rev.02, 05/2021
the duration of measurement in seconds. The higher you set the measure-
ment time, the more accurate the result.
4.
Click
<Ok>
to accept the value.
5.
Click on the button
<Perform adjust now>
(Fig. 87, item 1)
.
A new window opens (Standard Adjust Count Rate).
6.
Click on the button
<Start>
(Fig. 87, item 5)
to start the measurement.
7.
Click
<Ok>
(Fig. 87, item 4)
to accept the values.
The standard adjust was performed.
IMPORTANT
When executing an Adjust, the measurement parameter set was overwritten.
If the system should be recalibrated, a recall of the measurement parameter
set to the calibration parameter must be performed (see subchapter “re-
call”. Otherwise the adjust will become lost.