5 Electric Installation
LB 470RID Level
44
56925-1BA2 Rev.02, 05/2021
5.1.1
Circuit Breaker
A circuit breaker according to DIN EN 61010-1
•
must be available,
•
must be easily accessible for the maintenance personnel and
•
is to be included in the company-internal documentation.
The master EVUs / slave modules are not equipped with a separate ON/OFF switch
to connect or disconnect the voltage supply. Make sure that the system can be de-
energized via the external power supply.
The circuit breaker can be installed as an automatic fuse or switch and has to com-
ply with the requirements according to IEC 947-1 and IEC 947-3. If a fuse is applied,
it must not be triggered under a current strength of 4 A per device.
IMPORTANT
The circuit breaker must be located near the device and be properly marked as
belonging to it.
5.1.2
Cables and Lines
Lines are to be connected with special care.
Connection lines and routing must comply with the applicable regulations.
When routing the cables, make sure that the cable insulation cannot be me-
chanically damaged by sharp edges or movable metal parts.
Use the approved Berthold cable or a cable with equivalent specifications for
the connection.
The detector is connected via a 2-core (0.2 ... 2.5 mm²) cable with approx. 5 ... 10
mm diameter. A screened cable must be used in systems with extremely strong
electrical noise. The screen may only be laid out on one side of the detector. The
maximum cable length depends on the cable resistance, which may not exceed a
total (there and back) of 40 ohms. For standard cables from Berthold (Mat. No.
32024), this results in a cable length of 1000m, from the evaluation unit to the
detector.
When routing the connection lines, make sure that
signal lines (detector cables, power signal cables) are not laid together with
supply lines,
no dirt or moisture reaches the connection room,
the conductors are not damaged when the cable insulation is removed,
the conductor insulation or the sleeve of the wire end ferrules reach into the
housing of the terminal unit,
blank, conductive segments of the lines (e.g. wires of a litz wire) do not reach
outside the terminal unit,
the wire end ferrule or the stripped wire have a length of 8 mm so that the
wire is held securely in the clamp,
the line insulation reaches into the sleeve of the wire end ferrule if these com-
ponents are used,
the admissible minimum bending radius for the respective line cross-section is
not exceeded and
the cables are laid out in a strain-relieved and friction-free manner.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...