9 Troubleshooting
LB 470RID Level
184
56925-1BA2 Rev.02, 05/2021
9.2.4
Detector
Code
Message
NAMUR107 Help Text
M501
Detector not found
F
Lost connection to at least one detector.
Check detector settings and connections.
It is recommend to have the device
checked by Berthold Service, even if it
seems to work normally.
M502
Detector comm. error
M
Temporarily lost connection to at least
one detector. Check detector settings
and connections. Contact Berthold ser-
vice, if this event occurs repeatedly.
M503
Detector failure
F
At least one detector registers “Failure".
Check detector events.
M504
Detector out of spec.
S
At least one detector registers "out of
specification". Check detector events.
M505
Detector function check
C
At least one detector registers "function
check". Check detector events.
M506
Detector maintenance
M
At least one detector registers "mainte-
nance". Check detector events.
M599
Internal program error
F
Internal system failure. Powercycle de-
vice. If the event remains it could be
caused by a defective hardware. Contact
Berthold service.
9.2.5
RS 485 Interface
Code
Message
NAMUR107 Help Text
M699
Internal program error
F
Internal software failure. Restart the de-
vice. Contact Berthold service, if this
event occurs repeatedly.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...