7 Main Menu Device Setup
LB 470RID Level
92
56925-1BA2 Rev.02, 05/2021
Perform Plateau Measurement
Device Setup | Setup | Sensors | [NAME DETECTOR] | Plateau | Plateau Measurement
IMPORTANT
The environmental conditions and the dose rate must be constant during the
plateau recording.
Observe the operating manual of the detector.
1
2
3
4
5
6
Plateau curve
Button
<Start>
Window “Record plateau”
Button
<Yes>
Button
<No>
Event message code
7
8
9
10
11
12
Notice Maintenance
Notice “DET”
Progress bar measuring process
Progress bar measuring step
Button
<Stop>
New plateau curve
Fig. 61
Recording a plateau curve
1.
Click on
<Start>
(Fig. 61, item 2) to perform a plateau measurement.
The confirmation message "Record plateau" (Fig. 61, item 3) opens.
2.
Confirm with
<Yes>
(Fig. 61, item 4).
The EVU switches to mode "DET" (Fig. 61, item 8) and the current measure-
ment is stopped.
The information (Fig. 61, item 6 - 8) from the plateau measurement are dis-
played in the status information. The LED Run flashes on the EVU during the
plateau measurement. The LED “Warning” LED lights up at the same time.
If you click on the
< Stop >
button during the measurement, the measuring
process is interrupted. The measurement data are invalid and will be deleted.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...