LB 470RID Level
4 Installation
56925-1BA2 Rev.02, 05/2021
37
4.5
Installation in the Wall Housing
The wall housing may be equipped differently, depending on requirements (refer
to document “Technical Information”). To do this, a corresponding terminal panel
is located in the wall housing.
NOTICE
The master EVUs / slave modules must be secured against pulling out by fix-
ing screws (Fig. 14, item 4).
The device must be disconnected from the mains voltage before it is pulled
out.
Installation of the modules (master-slave)
1
2
3
4
Master-slave connector
Socket board
Guide rails
Fixing screws
Fig. 14 Installation of the modules (master-slave)
1.
Set modules into the guide rails (Fig. 14, item 3) and push it gently until the
plug connector of the module (Fig. 14, item 2) is inserted into the socket
board.
2.
Tighten all fixing screws (Fig. 14, item 4).
The modules are installed correctly.
3.
Unused slots must be covered with dummy panels.