4 Installation
LB 470RID Level
40
56925-1BA2 Rev.02, 05/2021
4.6.1
Installation with Clamp Blocks
j
DANGER
Danger to life from electric shock!
Installation/maintenance may only be carried out if the device has been de-
energized.
Test of absence of harmful voltages when the front side is open.
In case of an electric shock, carry out first aid measures and immediately call an
emergency service.
NOTICE
The master's EVUs / slave modules must be secured against pulling out by
fixing screws (Fig. 16, item 6).
The device must be disconnected from the mains voltage before it is pulled
out.
1
2
3
Master EVU
Slave module
Clamp block
4
5
6
Drilling
Guide rail
Fixing screw
Fig. 16
19" subrack with clamp block (Ex: 1x Master, 9x Slave)
1.
Set master EVU / slave module (Fig. 16, item 1, item 2) in the guide rails (Fig.
16, item 5).
2.
Carefully slide module into the subrack until the plug connector is inserted
into the clamp block.
3.
Tighten fixing screws (Fig. 16, item 6).
4.
Unused slots must be covered with dummy panels.
The EVU is correctly inserted and can be connected.