55
Class I, IPLeak, alternative leakage current measurement
L, N and PE conductors are bridged
Connection diagram
VDE 0751-1 (applied parts of type F)
Circuit diagram
VDE 0751-1 (applied parts of type F)
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The testing voltage is applied and must be tested in all switch
positions (device functions).
Use the test probes to scan all accessible conductive parts and
applied parts.
Class II, IPLeak, direct current measuring method
The polarity of L-N must be reversed.
Connection diagram
VDE 0751-1 (applied parts of type F)
Circuit diagram
VDE 0751-1 (applied parts of type F)
Place the test sample onto an insulated surface.
Insert the shock-proof plug of the test sample into the test socket
of the appliance tester.
The test sample is supplied with mains voltage and must be
tested in all switch positions (device functions).
Use the test probes to scan all accessible conductive parts and
applied parts.
Class II, IPLeak, alternative leakage current measurement
L and N conductors are bridged
Connection diagram
VDE 0751-1 (applied parts of type F)
Circuit diagram
VDE 0751-1 (applied parts of type F)
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The testing voltage is applied and must be tested in all switch
positions (device functions).
Use the test probes to scan all accessible conductive parts and
applied parts.
Measurement
Make sure that the test sample is free of mains voltage.
Connect the mains plug of the test sample to the test socket of
the appliance tester.
Select a measuring
method.
Select an applied part.
Direct current measuring
method
Alternative leakage current
measurement
Tap the “
Start
” button.
Tap the “
Start
” button.
Use the test probe to scan all
of the active applied parts of
the test sample.
Use the test probe to scan all
of the active applied parts of
the test sample.
Tap the “
L<->N
” button.
Use the test probe to scan all
of the active applied parts of
the test sample.
Tap the “Stop” button to stop the test.
The “Stop” button must only be tapped if the test time for
automatic testing is set to ‘infinite’ (0).
Otherwise, the test will be stopped automatically.
The “
Test result
” screen opens.