Page 1-42
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Specifications
Chapter 1: Introduction
Table 1-12. Cabling Requirements from the Master Alarm Module to the
Source Equipment
Table 1-13. Network Communication Cable
Table 1-14. Audible Specifications
Table 1-15. Environmental Specifications
Environmental Specification
Dimensions
Operating temperature
50°F to 120°F (10°C to 49°C)
Storage temperature
- 40°F to 185°F (- 40°C to 85°C)
Operating humidity
0% to 85% non-condensing
Storage humidity
0% to 85% non-condensing
FCC Part 15
Class A, emissions
Audible Specifications
Dimensions
Area Alarm audible output
80 dBA at one meter minimum
Master Alarm audible output
80 dBA at one meter minimum
Local Alarm audible output
80 dBA at one meter minimum
Specifications
Dimensions
Master Alarm to alarm network devices
Standard Ethernet 10BaseT cable
Network Communications Board to alarm
network devices
Standard Ethernet 10BaseT cable
Specifications
Dimensions
Distance
Maximum 10,000' (3048 m)
Cable
Stranded, 22-gauge, shielded, twisted pair
Signal Power
20 V DC, 20 mA maximum