No Display on One Area Alarm Module
Chapter 2: Troubleshooting Procedures
2.27 No Display on One Area Alarm Module
1. The displays on the other Area Alarm Modules are ON.
Yes
No
L J
Go to RAP 2.24 on page 2-30. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 2.
2. The ribbon cable from the Area Alarm Connector P.C. Board to the Area
Alarm Module is connected.
Yes
No
L J
Connect the ribbon cable from the Area Alarm Connector P.C.
Board to the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 3.
3. Replace the ribbon cable (refer to procedure 4.19 on page 4-48). The problem
is solved.
Yes
No
L J
Call BeaconMedæs Technical Support.
4. Go to “Final Actions” on page 2-7.
2.28 Audible Signal Does Not Silence
1. Pressing the
Silence key
cancels the audible signal.
Yes
No
L J
Replace either the Master Alarm Module (refer to procedure 4.2
on page 4-10) or the Area Alarm Module (refer to procedure 4.9
on page 4-26) on the alarm panel that will not silence. If this
solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call BeaconMedæs Technical Support.
2. Go to “Final Actions” on page 2-7.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...