Page 4-22
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Area Alarm Breakout P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.7
Area Alarm Breakout P.C. Board
Tools required:
Phillips head screwdriver
5/32" hex wrench
Needle nose pliers
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-8 on page
4–22).
Figure 4-8. Area Alarm Breakout P.C. Board
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (H) to the OFF position.
3. Push open the ears of the connectors (E) to remove the ribbon cable (B).
4. Note the positions of the wiring on the connector (I) for the transducers.