Page 4-50
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Remote-Mount Transducer and Demand Flow Valve Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.20 Remote-Mount Transducer and Demand Flow Valve Assembly
Tools required:
Adjustable wrench
Phillips head screwdriver
Small screwdriver
Leak detection solution
Teflon® tape
Removal
WARNING:
Before making or breaking any gas or vacuum line connections, ensure
that the system is depressurized to avoid possible injury.
1. Ensure that the gas or vacuum line is not under pressure.
WARNING:
Do not remove the threaded insert from the transducer. Doing so may
result in attaching the transducer to the incorrect gas or vacuum type,
endangering patient welfare.
2. Leave the threaded insert (A) in the transducer (B). The transducer is not a
serviceable item (see figure 4-21 on page 4–51).
3. Using an adjustable wrench, remove the demand flow valve (C) from the tee
port (D).
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...