While in the configuration mode, the front panel keys operate differently than
when in normal operation mode. Refer to the following:
1. Scroll through the options, by pressing the
Increment key
or
Decrement key
(see figure 1-59 on page 1–117).
Figure 1-59. Area Alarm Module
2. When the code for the option that you need shows on the
display
, stop scroll-
ing.
3. View the setting for that option by pressing and holding the
Test key
.
4. To change the setting, do the following:
a. Press and hold the Test key while scrolling through the available options
with either the Increment key or Decrement key.
b. When the intended setting displays, release the
Test key
.
NOTE:
If nA displays, that option is not available.
5. You can return to normal operation and save the changes that you have made
at any time.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-117
Configuring the Area Alarm by Using the Front Panel
Chapter 1: Introduction
1
Display
Silence Key
Test Key
Increment key
Decrement Key