Page 1-118
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring the Area Alarm by Using the Front Panel
Chapter 1: Introduction
Returning to Normal Operating Mode
1. To return to normal operation and save any changes that you have made, do
one of the following:
•
Press the configuration button (CE) or (CF) (see figure 1-58 on page
1–116).
•
Turn the power supply switch (DR) to the OFF position, and then back
to the ON position (see figure 1-57 on page 1–115).
If you do not press any keys for 10 minutes, the alarm returns to normal opera-
tion and saves any changes that you have made.
2. While
GASt
is being displayed, you can view the type of gas that the Area
Alarm Module is monitoring by pressing the
Increment key
or the
Decrement key
. For a list of available gases, refer to table 1-31 on page 1-
118.
NOTE:
The type of gas or vacuum being monitored cannot be changed.
Table 1-31. Area Alarm Module Gas Types
Display Shows
Gas type
oy9n
Oxygen
M7Ed
Medical air
UAcu
Vacuum
WJA9
WAGD (waste anesthetic gas disposal)
nito
Nitrous oxide
nit9
Nitrogen
cArb
Carbon dioxide
hELi
Helium
Oycd
Oxygen and carbon dioxide mixture
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...