Getting Started
WARNING:
Only facility-authorized personnel should troubleshoot the MedPlus
TotalAlert® Alarm Network. Troubleshooting by unauthorized personnel
could result in personal injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the nor-
mal operation of the product can be confirmed by answering Yes or No to the
statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one compo-
nent is listed, replace them in the given order.
To begin gathering information about the problem, start with
Initial Actions
.
To verify the repair after completing each corrective action (replacing or adjust-
ing a part, seating a connector, etc.), perform the
Function Checks
.
To verify the repair, perform the
Final Actions
after the Function Checks.
If the troubleshooting procedures do not isolate the problem, call
BeaconMedæs Technical Support at 1-888-4MEDGAS (1-888-463-3427) for
assistance.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-3
Getting Started
Chapter 2: Troubleshooting Procedures
2
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...