Page 4-36
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Local Alarm Transformer
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.13 Local Alarm Transformer
Tools required:
Phillips head screwdriver
Needle nose pliers
Removal
Shock Hazard:
Turn off the source power to the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
1. Turn off power to the alarm at the facility’s source.
2. Using a phillips head screwdriver, loosen the four screws (A) enough to
allow the front panel (B) to slide off (see figure 4-14 on page 4–36).
Figure 4-14. Local Alarm Transformer
3. Slide the front panel (B) up and off of the back box (C).
4. Unplug the buzzer wiring (D).