3-30 | IPS-4-UV Analyzer-GPDiv2
Adjusting the Zero/Span Fluid Flow Rate
When the zero or span (if used) flow solenoid is in an energized state,
Instrument Air (or Nitrogen or distilled water) flows into the sample path
via a solenoid manifold block and a flow meter. The flow rate of Zero/
Span (if used) fluid must be high enough to adequately flush the Measur-
ing Cell of sample fluid to obtain a “good zero.”
Typically, the IPS-4 Analyzer controls the zero (and span, if used)
flow rate; no adjustments are required.
If the customer provides the sample system zero, run the manual zero
at a flow rate of 2.5 L/minute (0.08 scfm). Since excess air flow into
the analyzer sample system will cool the Measuring Cell and Oven/
Sample System Enclosure, the optimum Zero fluid flow rate is the
lowest consistent with a good zero.
A good zero can be defined as a state where the concentration outputs
are stable and further increases in Zero fluid flow rate do not reduce the
concentrations observed on the analyzer User Interface Panel. A mid-scale
flow indicator (or meter) should be sufficient for this purpose.
While the sample system is warming up, or when the analyzer is in
standby mode (
Not Ready
state), zero flow may be desired. During these
conditions, it is important that the sample system is kept free of contami-
nation by the sample fluid.
For additional information, refer to the Sample System
Manual Supplement
in the “Supplemental Information” tab of this manual or in the Documen-
tation Package shipped with the analyzer.
Adjusting the Sample Flow Rate
Applications With an Aspirator:
The Aspirator Pressure Regulator is used to control the sample flow
through the system. A typical sample flow rate is in the range of
3.0–5.0 L/minute (0.1–0.2 scfm) and can be achieved by adjusting the Aspi-
rator Pressure higher than 70 kPa (10 psi) above the sample stream pres-
sure. View the flow indicator (or meter), if used.
To avoid damaging the pressure transducer, do not set the pressure
higher than 105 kPag (15 psig).
NOTE
NOTE
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...