4-30 | IPS-4-UV Analyzer-GPDiv2
System Settings Screen
The System Settings screen (Figure 4-30) allows the user to view and/or
setup analyzer system parameters.
Figure 4-30.
System Settings
Screen.
> Log in/Set Passwords
Backup/Restore
Ethernet
Network
System
Customization
HOME ALARMS
STATUS SETUP
Login/Set Passwords Screen
The
Login/Set Password
screen (Figure 4-31) allows you to set up pass-
word protection for users who need access to modify analyzer configura-
tion parameter settings, or restrict users to read-only function. The ana-
lyzer is shipped from the factory without a password. Upon powering up
the analyzer for the first time, all configurable parameters can be changed
until a password is entered in the
Change Password
function.
Figure 4-31.
Login/Set Password
screen.
Enter Password
Change
Password
Lock
HOME ALARMS
STATUS SETUP
When you first power up the analyzer, or following a system reset, the
software security level defaults to read-only mode – you will have to
re-enter the password to allow changes. Password protection can also be
removed, as described under “Removing the Password.”
When a password has been entered, the software will remain in Edit
(read-write) mode as log as the user is working from the User Interface
Panel. If no keys are pressed within 10 minutes, the software will revert to
read-only mode. The password will have to be re-entered before further
changes can be made.
Enter Password
Use this function to enter a password that will allow you to make
changes to many of the analyzer configuration settings, or to perform
calibration and other functions. Because the analyzer is shipped from
the factory without a password, you must first enter a password in the
Change Password
function.
Record and store the password in a safe
place for reference.
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...