4-20 | IPS-4-UV Analyzer-GPDiv2
Calibration Actions Screen
The
Calibration Actions
screen displays all of the available analyzer cali-
bration functions and allows the user to initialize the calibration functions.
Figures 4-17a and 4-17b illustrate the differences between analyzers that
use single-species spanning and those that use multi-species spanning.
Figure 4-17a.
Calibration screen,
applications with
single span fl uid.
Start Zero Calibration
Start Span Calibration
Start Zero Cal+Span Calibration
Inject Zero Fluid
Inject
Span
fl uid
Start
measure
HOME ALARMS
STATUS SETUP
Figure 4-17b.
Calibration screen,
applications with
multi-span fl uids.
Start zero calibration
Select Span Fluid
Start span Calibration
Start Zero Cal + Span Calibration
Inject
Zero
fl uid
Inject
Span
fl uid
HOME ALARMS
STATUS SETUP
Start zero calibration
This function forces the analyzer into the Zero calibration portion of
the normal operating sequence, starting with the “
Zero Flush
” mode,
then “
Zero
.” The analyzer zero is automatically adjusted based on the
average reading during the zero calibration.
Select Span Fluid (for applications that require multiple span fluids)
This function is used in applications where multiple span fluids are
required. The span fluid connected to the
Cal (Span) Fluid Inlet
port
must correspond to the fluid selected from the options listed under
this parameter.
To select a span fluid, press
ENTER
to go into edit mode. Use the
UP
/
DOWN
arrow key to select the fluid you wish to span and then press
ENTER
to select the span fluid. Next, select
Start span calibration
and
press
ENTER
to start the span.
Start span calibration
This function forces the analyzer into the Span mode of the normal
operating sequence, starting with the “
Span Gas Flush
” mode. The
analyzer span (calibration) is automatically adjusted based on an inte-
grated reading during the span calibration. After the span is complete,
the analyzer will return to normal operation (
Measure
mode).
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...