Supplemental Information
|
G-3
119
19 Vent temperature high limit
alarm
Alarm
Bench temperature (283) is outside
of a band set by parameter 278
around the setpoint (284)
119
20 Vent temperature low limit
alarm
Alarm
119
21-
32
Not
used
120
Not
used
121
Not
used
122
Not
used
123
Not
used
The bit map of System Warning (parameter 514, 515, 516; 32-bit integer, modbus
register 1034, 1035, 1036, 1037, 1038, 1039) contains warnings as defined:
Modbus
register
Bit
Description
Type
Condition
103
4
1-
30
Not
used
1034
31
Span calibration is required
Warning Detected when span calibration was
never performed prior
1034
32
Zero calibration is required
Warning Detected when zero calibration was
never performed prior
1035
1 Zero
calibration
failure
Warning Unable to perform zero calibration
1035
2
Span 1 calibration failure
Warning Unable to perform span calibration for
analyte 1
1035
3
Span 2 calibration failure
Warning Unable to perform zero calibration for
analyte 2
1035
4
Span 3 calibration failure
Warning Unable to perform zero calibration for
analyte 3
1035
5
Span 4 calibration failure
Warning Unable to perform zero calibration for
analyte 4
1035
6
Span 5 calibration failure
Warning Unable to perform zero calibration for
analyte 5
1035
7
Span 6 calibration failure
Warning Unable to perform zero calibration for
analyte 6
1035
8
Span 7 calibration failure
Warning Unable to perform zero calibration for
analyte 7
1035
9
Span 8 calibration failure
Warning Unable to perform zero calibration for
analyte 8
1035
10
Dark read failure
Warning Dark pixel readings are not within a
range of pre-defined values
1035
11 Battery failure
Warning System battery fails
1035
12 Non-volatile
memory
checksum failure
Warning Non-volatile memory, where system
configuration is stored, does not return
correct checksum value
1035
13
Zero drift exceeded
Warning Zero drift has exceeded a pre-defined
allowable value
1035
14
Zero average intensity
failure
Warning Zero average readings of the intensity
of the diode array are out of
specification
1035
15 Electronic
check
failure
Warning Verification of the electronic bench fails
1035
16
Verification operation failure Warning Fail to check verification filter
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...