Overview | 1-7
Four independent analog output signal channels
The full-scale concentration for each output are configurable as either
0–20 mA or 4–20 mA.
Sample control system that can control and monitor the state of up to five tem-
perature-control zones within the sample system
Temperature-control zones include the Oven/Sample System Enclo-
sure and Optical Bench. Three other zones are currently spare. Sample
and Vent Lines can be monitored only.
The temperature-control zones are protected by independent over-
temperature circuits. If any of these zones are found to be out of
range, the analyzer automatically forces its sample system into
Not Ready
mode to avoid contamination of the system.
User Interface Panel
Provides access to a readout of system parameters, alarm set points,
local diagnostics, and user input. The calibration and configuration
functions can be protected by a password.
Up to eight status relays
Indicate faults, non-critical conditions, and operational status of the
analyzer.
Reliable electronics control system
Controls the overall operation of the analyzer.
Ethernet Communication
Allows for remote communication with the analyzer for early-detec-
tion diagnostics (troubleshooting) to determine imminent failure of
parts (e.g., lamp failure, due to diminishing voltage levels).
Gas and Liquid applications
Sample systems designed to accommodate gas and liquid applications.
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...