Specifi cations | 2-3
indicate the operational status of the analyzer. Each relay provides one
SPST (Form A) dry (potential free) contact.
RS485 serial port, isolated.
RS232 serial port, non-isolated.
Fast Ethernet (IEE802.3).
Modbus TCP (Supports single Modbus client).
Sample System Limits
Sample Pressure
Up to 10,000 kPag (1,450 psig / 100 barg) can be accommodated, depend-
ing on the application. See product markings or customer-specific draw-
ings in the “Supplemental Information” tab of this manual.
Oven/Sample System Enclosure Temperature
For applications with a heated sample system, the Oven will operate at
temperatures up to 150 °C (302 °F).
Sample Transport
Gas Applications:
By aspiration, using Instrument Air or N
2
as the drive gas.
Liquid Applications:
Using sample stream pressure.
Minimum 5psi differential pressure inlet to outlet.
Instrument Air Requirements
The aspirator medium must be transported in 316 stainless steel.
Minimum pressure:
490–700 kPag (70–100 psig / 4.8–6.9 barg)
Air Quality:
As per ANSI/ISA-S7.0.01 (1996) Quality Standard for Instrument Air.
Summary of Contents for IPS-4-UV
Page 15: ...Contents xv...
Page 16: ...xvi IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 24: ...1 8 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 66: ...3 36 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 144: ...6 6 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 180: ...A 2 IPS 4 UV Analyzer GPDiv2 Analyzer Light Path Schematic...
Page 183: ...Appendix Drawings A 5 Detector Interface Board 100 1937...
Page 184: ...A 6 IPS 4 UV Analyzer GPDiv2 Detector Board 100 1955...
Page 185: ...Appendix Drawings A 7 Display Interface Board 100 1978...
Page 186: ...A 8 IPS 4 UV Analyzer GPDiv2 Customer Connection Board 100 1971...
Page 187: ...Appendix Drawings A 9 Relay Board 100 1942...
Page 188: ...A 10 IPS 4 UV Analyzer GPDiv2 Analog Board 100 1922...
Page 189: ...Appendix Drawings A 11 MCU Board 100 1914...
Page 190: ...A 12 IPS 4 UV Analyzer GPDiv2 Xenon Lamp Board 100 2061...
Page 191: ...Appendix Drawings A 13 Electronics Enclosure Wiring Diagram Sheet 1 of 3 WX 102877...
Page 192: ...A 14 IPS 4 UV Analyzer GPDiv2 Electronics Enclosure Wiring Diagram Sheet 2 of 3 WX 102877...
Page 193: ...Appendix Drawings A 15 Electronics Enclosure Wiring Diagram Sheet 3 of 3 WX 102877...
Page 194: ...A 16 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...
Page 196: ...S 2 IPS 4 UV Analyzer GPDiv2 This page intentionally left blank...