4.6.5 Torque Calibration Routine
Used to calibrate tool using a Master Transducer. The following are
steps to calibrate tool.
1. Press the “Start Calibration” button.
2. Run Tool to Final Torque.
3. Enter external transducer (Master) value in Measured Torque box.
4. Press “Send New Cal Values to Tool” button.
5. New Cal value is calculated automatically and assigned to tool ID
Board.