9-5
Clock Recovery Commands
CRATe
CRATe
Command
(83496A)
:CRECovery{1 | 3}:CRATe <data_rate>
This 83496A command sets or queries an 83496A module’s data rate setting. Although the
command
can be used with 83496A modules, use the preferred CRATe
command in all new programs controlling an 83496A module. The data rate for standard
83496A modules ranges from 50 Mb/s to 7.10 Gb/s. The data rate for Option 200 modules
ranges from 50 Mb/s to 13.50 Gb/s. The data rate can be specified to 6 significant digits. The
default setting is 2.488 Gb/s.
Restrictions
Software revision A.04.20 and above.
Example
10 OUTPUT 707; ":CRECOVERY1:CRATE 4.25E9"
Query
:CRECovery{1 | 3}:CRATe?
Returned Format
[:CRECovery{1 | 3}:CRATe <data_rate><NL>
INPut
Command
(83495/6A)
:CRECovery{1 | 3}:INPut{ELECtrical | OPTical | DIFFerential | EINVerted}
Selects the clock recovery input on 83495A and 83496A modules. On 83495A modules, OPTi-
cal is the default setting. On 83496A modules, ELECtrical is the default setting. The argu-
ments, DIFFerential and EINVerted (electrical inverted), are available on 83496A modules
only.
Restrictions
83495A and 83496A modules. Software revision A.03.10 and above for 83495A module. Soft-
ware revision A.04.20 and above for support of 83496A modules.
Example
10 OUTPUT 707;":CRECOVERY1:INPUT ELECTRICAL"
Query
:CRECovery{1 | 3}:INPut?
Returned Format
[:CRECovery{1 | 3}:INPut] {ELECtrical | OPTical | DIFFerential | EINVerted}<NL>
LBANdwidth
Command
(83495/6A)
:CRECovery{1 | 3}:LBANdwidth {BW270KHZ | BW300KHZ | BW1500KHZ | BW4MHZ | CONTinuous}
Sets the loop bandwidth on 83495A and 83496A modules to a value as listed in
. The default setting is 300 kHz for 83495A modules and 270 kHz for 83496A mod-
ules. The CONTinuous argument (83496A Option 300 only) can be returned in queries but
can not be sent in a command string. CONTinuous is returned whenever the loop bandwidth
of an 83496A Option 300 module is set to a value other than the LBANdwidth standard val-
ues. When the CONTinuous argument is returned, use the CLBandwidth command to query
the actual value.
Refer to “CLBandwidth” on page 9-4
.
Do not use this command with 83496A Option 300 modules. Instead, use the command
.
Restrictions
83495A modules , software revision A.03.10 and above. 83496A modules (except Option
300), software revision A.04.20 and above.
Example
10 OUTPUT 707;":CRECOVERY1:LBANDWIDTH BW4MHZ"
Query
:CRECovery{1 | 3}:LBANDWIDTH?
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......