9-7
Clock Recovery Commands
ODRatio
Example
10 OUTPUT 707;":CRECOVERY1:LOCKED?"
ODRatio
Command
(83496A)
:CRECovery{1 | 3}:ODRatio <divide_ratio>
This 83496A command sets or queries the output clock divide ratio. This determines the data
rate at the front-panel recovered clock output. The ratio can be set to a value of 1, 2, 4, 8, or
16. Sending this command while the output divider is set to auto (
), results in a settings conflict error.
Restrictions
83496A. Software revision A.04.20 and above.
Example
10 OUTPUT 707; ":CRECOVERY1:ODRATIO 2"
Query
:CRECovery{1 | 3}:ODRatio?
Returned Format
[
:CRECovery{1 | 3}:ODRatio] <divide_ratio><NL>
ODRatio:AUTO
Command
(83496A)
:CRECovery{1 | 3}:ODRatio:AUTO {{ON | 1} | {OFF | 0}}
This 83496A command enables or disables the module’s capability to automatically set the
divide ratio for the front-panel recovered clock output. With auto on, the instrument auto-
matically selects an output divide ratio setting to 1:1 for frequencies equal to or less than 7.1
GHz or 1:2 for frequencies greater than 7.1 GHz.
Restrictions
83496A. Software revision A.04.20 and above.
Example
10 OUTPUT 707; ":CRECOVERY1:ODRATIO:AUTO ON"
Query
:CRECovery{1 | 3}:ODRatio:AUTO?
Returned Format
[
:CRECovery{1 | 3}:ODRatio:AUTO] {{ON | 1} | {OFF | 0}}<NL>
RATE
Command
(83491/2/3/4/5/6A)
:CRECovery{1 | 3}:RATE {TOData | R155 | R622 | R1062 | R1250 | R2125 | R2488 | R2500 | R2666 | R9953 | R10312
| R10664 | R10709 | RANGE10G}
This command sets the clock recovery module’s data rate. The available rates for each mod-
ule, with associated command arguments, are listed in
. Rate parame-
ters are nominal and reflect front-panel labels and not actual data rates. The TOData
argument selects triggering on the data. Although this command will work with 83496A mod-
ules, on new programs for the 83496A module, use the command
Restrictions
The CONTinous query response is only returned by 83496A modules and requires software
revision 4.20 and above.
Example
10 OUTPUT 707;":CRECOVERY3:RATE R2488"
Query
:CRECovery{1 | 3}:RATE?
Returned Format
The CONTinuous query response appears in queries only and can not be sent in a command
string. CONTinuous is returned whenever the data rate of an 83496A module is not one of the
standard values set using the CRECovery:RATE command. If the CONTinuous argument is
returned, use the CRECovery:CRATE command to query the actual value.
.
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......