21-3
TDR/TDT Commands (Rev. A.05.00 and Below)
NVALid?
Query
:TDR{2 | 4}:HPOLarity?
Returned Format
[:TDR{2 | 4}:HPOLarity] {POSitive | NEGative}<NL>
NVALid?
Query
:TDR{2 | 4}:NVALid?
Queries the specified TDR module to determine if valid normalization data exists. A 1 is
returned, if a valid normalization exists. Otherwise, a 0 is returned.
Restrictions
Software revision A.04.20 and A.05.00. TDR mode.
Example
10 OUTPUT 707;":TDR2:NVALid?"
Returned Format
[:TDR{2 | 4}:NVALid] {1 | 0}<NL>
Returned Format
[:TDR{2 | 4}:RESPonse<N>] {1 | 0}<NL>
PRESet
Command
:TDR{2 | 4}:PRESet
This command performs an automatic set up of the instrument for TDR or TDT measure-
ments, based on the stimulus. This command does the following:
• Turn on TDR channels.
• If the stimulus is set to EXT ernal (
), turn off channel 1 or 3 and
turn on channel 2 or 4.
• If the TDT destinations are not shown, turn on the TDT destination channels. (
• Set the timebase to 500 ps/div and positions the incident edge on screen.
• Turn on averaging and set best flatness (
see “Acquire Commands” in chapter 6
).
• For all channels that are on:
• Set the attenuation units to ratio.
• Set the attenuation to 1:1.
• Set the bandwidth to low (12.4 GHz). (Set high for external stimulus.)
• Set the units to volts.
• Set the channel scale to 100 mV/div.
• Set the channel offset to 200 mV or –200 mV for differential stimulus.
Restrictions
Software revision A.05.00 and below. TDR mode.
Example
The following example presets the instrument for TDR/TDT operations.
10 OUTPUT 707;":TDR2:PRESET"
RATE
Command
:TDR{2 | 4}:RATE {AUTO | <rate>}
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......