18-46
Measure Commands
VAMPlitude
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASURE:TVOLT? -.250,+3”
VAMPlitude
Command
:MEASure:VAMPlitude [<source>]
Calculates the difference between the top and base voltage of the specified source. Sources
are specified with the MEASure:SOURce command or with the optional parameter following
the VAMPlitude command.
<source> is the {CHANnel<N> | FUNCtion<N> | RESPonse<N> | WMEMory<N>} where <N>
is 1, 2, 3, or 4.
Mode
Oscilloscope and TDR modes.
Query
:MEASure:VAMPlitude? [<source>]
The query returns the calculated difference between the top and base voltage of the specified
source.
Returned Format
[:MEASure:VAMPlitude] <value>[,<result_state>]<NL>
<value> is the cCalculated difference between the top and base voltage. If SENDvalid is ON,
the <result_state> is returned with the measurement result. Refer to
for a list of the result states.
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;":MEASURE:VAMPLITUDE?2"
VAVerage
Command
:MEASure:VAVerage {CYCLe | DISPlay} [,<source>]
Calculates the average voltage over the displayed waveform. The source is specified with the
MEASure:SOURce command or with the optional parameter following the VAVerage com-
mand. The CYCLe parameter specifies to measure the average voltage across the first period
of the display. This option is valid in oscilloscope mode only. The DISPlay parameter specifies
to measure all the data on the display. This option is valid in both oscilloscope and TDR
modes. The <source> is {CHANnel<N> | FUNCtion<N> | WMEMory<N> | RESPonse<N>}
where <N> is an integer, from 1 through 4.
Mode
Oscilloscope and TDR (DISPlay option only) modes.
Query
:MEASure:VAVerage? {CYCLe | DISPlay}, [<source>]
The query returns the calculated average voltage of the specified source.
Returned Format
[:MEASure:VAVerage] <value> [,<result_state>]<NL>
<value> is the calculated average voltage. If SENDVALID is ON, the <result_state> is
returned with the measurement result. Refer to
result states.
Example
10 OUTPUT 707;”:SYSTEM:HEADER OFF”
20 OUTPUT 707;”:MEASURE:VAVERAGE? DISPLAY”
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......