24-6
Trigger Commands
SLOPe
SLOPe
Command
:TRIGger:SLOPe {POSitive | NEGative}
This command specifies the slope of the edge on which to trigger.
Query
:TRIGger:SLOPe?
The query returns the slope for the trigger.
Returned Format
[:TRIGger:SLOPe] {POSitive | NEGative}<NL>
Example
10 OUTPUT 707; ":TRIGger:SLOPe POSitive"
SOURce
Command
:TRIGger:SOURce {FPANel | FRUN | LMODule | RMODule}
This command selects the trigger input. Front Panel, Left Module, and Right Module are
inputs from the front panel of the instrument. Free Run is internally generated, and is not
affected by the settings of gates, level, slope, bandwidth, or hysteresis.
<trigger>
Front PANel, Left MODule, and Right MODule are inputs on the front of the instrument. Fre-
eRUN is internally generated and is unaffected by the settings for gated, level, slope, band-
width or hysteresis.
Query
:TRIGger:SOURce?
The query returns the current trigger source of the current mode.
Returned Format
[:TRIGger:SOURce] <trigger><NL>
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......