1-45
Introduction
Error Messages
Acquisition
Settings
Acquisition (Averaging) controls are disabled in Jitter mode. The following commands pro-
duce a "Settings conflict" error when executed in Jitter mode.
ACQuire:
ACQuire:
ACQuire:
Histograms
Histograms are turned off when entering Jitter mode. The following commands produce a
"Control is set to default" error.
HISTogram:
Software
Skewing of
Channels
All skew adjustments are disabled in jitter mode. The following commands produce a "Set-
tings conflict" error when executed in Jitter mode.
CALibrate:SKEW
CALibrate:SKEW:AUTO
Error Messages
This chapter describes the error messages and how they are generated. The possible causes
for the generation of the error messages are also listed in
.
Error Queue
As errors are detected, they are placed in an error queue. This queue is first in, first out. If
the error queue overflows, the last error in the queue is replaced with error –350, “Queue
overflow.” Anytime the error queue overflows, the oldest errors remain in the queue, and the
most recent error is discarded. The length of the instrument's error queue is 30 (29 positions
for the error messages, and 1 position for the “Queue overflow” message). Reading an error
from the head of the queue removes that error from the queue, and opens a position at the
tail of the queue for a new error. When all errors have been read from the queue, subsequent
error queries return 0, “No error.”
The error queue is cleared when any of the following occur:
• the instrument is powered up,
• a *CLS command is sent,
• the last item from the queue is read, or
• the instrument is switched from talk only to addressed mode on the front panel.
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......