9-10
Clock Recovery Commands
TDENsity?
TDENsity?
Query
(83496A)
:CRECovery{1 | 3}:TDENsity?
Use this 83496A query with 83496A modules to return the calculated edge density of the data
signal. The edge density value is the ratio of bit transistions to bits and is returned as a num-
ber between zero and one. Changes in edge density can cause the module to lose lock. If the
edge density value is invalid, the string “
9.37
” is returned.
Restrictions
Software revision A.04.20 and above.
Example
10 OUTPUT 707;":CRECOVERY1:TDENSITY?"
Returned Format
[:CRECovery{1 | 3}:TDEN] <edge_density><NL>
Table 9-5. Signal Present Return Status vs. Receiver Number
Module Model
Receiver 1
Short Wavelength
Receiver 2
Long Wavelength
83491
0
0
83492
a
a. Only one receiver at a time can have a signal present.
1/0
1/0
83493
0
1/0
83494
0
1/0
83494 Option 103
0
1/0
83494 Option 106
0
1/0
83494 Option 107
0
1/0
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......