24-3
Trigger Commands
BWLimit
BWLimit
Command
:TRIGger:BWLimit {DIVided | HIGH | LOW}
This command controls an internal lowpass filter and a divider in the 86100A trigger. The
bandwidth of the trigger is limited to approximately 100 MHz. DIVided mode is unaffected by
the level, hysteresis, and slope settings. The DIVided parameter is only valid if the mainframe
has option 001.
Query
:TRIGger:BWLimit?
The query returns the current setting for the specified trigger input.
Returned Format
[:TRIGger:BWLimit] {HIGH | LOW| DIV}<NL>
Example
10 OUTPUT 707;”:TRIGGER:BWLIMIT LOW”
DCDRatio
Command
:TRIGger:DCDRatio <data_to_clock_divide_ratio>
This command is used to set the data-to-clock divide ratio used by pattern lock trigger mode.
<data_to_clock_divide_ratio> must be one of the following integers: 1, 2, 4, 5, 8, 10, 15, 16,
20, 25, 30, 32, 35, 40, 45, 50, 64, 66, 100, 128.
Restrictions
Software revision A.04.00 and above (86100C instruments)
Query
:TRIGger:DCDRatio?
This query returns the current setting of data-to-clock divide ratio.
Returned Format
[:TRIGger:DCDRatio] <data_to_clock_divide_ratio><NL>
Example
10 OUTPUT 707; "
:
TRIGger:DCDRatio 16"
DCDRatio:AUTodetect
Command
:
TRIGger:DCDRatio:AUTodetect {{ON | 1} | {OFF | 0}}
This command enables or disables automatic detection of the data-to-clock divide ratio.
When disabled, use the
:TRIGger:DCDRatio
command to set the data-to-clock divide ratio. When
enabled, use the
:TRIGger:PLOCk:AUTodetect
command to initiate automatic detection.
Restrictions
Software revision A.04.00 and above (86100C instruments)
Query
:TRIGger:DCDRatio:AUTodetect?
Returned Format
[
:TRIGger:DCDRatio:AUTodetect] {1 | 0}<NL>
Example
10 OUTPUT 707; ":TRIGger:DCDRatio:AUTodetect ON"
GATed
Command
:TRIGger:GATed {ON | 1 | OFF | 0}
This command enables or disables the ability of the instrument to respond to trigger inputs.
Query
:TRIGger:GATed?
The query returns the current gated setting.
Returned Format
[:TRIGger:GATed] {1 | 0}<NL>
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......