18-42
Measure Commands
SINTegrity:SIGNal
SINTegrity:SIGNal
Command
MEASure:SINTegrity:SIGNal {CLOCk | DATA}
This command specifies the type of signal being measured in Jitter and Signal Integrity mode.
It replaces the obsolete command :MEASure:JITTer:SIGNal.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Query
MEASure:SINTegrity:SIGNal?
Returned Format
[:MEASure:SINTegrity:SIGNal] {CLOCk | DATA}<NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:SIGNAL CLOCK”
SINTegrity:SIGNal:AUTodetect
Command
MEASure:SINTegrity:SIGNal:AUTodetect {ON | 1 | OFF | 0}
This command turns automatic detection of the signal type (clock or data) on or off. The
automatic detection happens during an autoscale. This command replaces the obsolete com-
mand :MEASure:JITTer:SIGNal:AUTodetect.
Restrictions
86100C (software revision A.07.00 and above) with Jitter Mode including Advanced Ampli-
tude Analysis/RIN/Q-Factor application.
Query
MEASure:SINTegrity:SIGNal:AUTodetect?
Returned Format
[:MEASure:SINTegrity:SIGNal:AUTodetect] {1 | 0}<NL>
Example
10 OUTPUT 707;”:MEASURE:SINTEGRITY:SIGNAL:AUTODETECT ON”
SOURce
Command
:MEASure:SOURce <source>[,<source>]
Selects the source for measurements. You can specify one or two sources with this command.
All measurements except MEASure: DEFine:DELTatime are made on the first specified
source. The delta time measurement uses two sources if two are specified. If only one source
is specified, the delta time measurement uses that source for both of its parameters. The
source is always color grade/gray scale data in eye mode, except for average optical power
and histogram measurements. This is a global definition. It is used for all subsequent remote
measurements unless a different source is specified with the optional source parameter in the
measure command.
<source> is {CHANnel<N> | FUNCtion<N> | RESPonse<N> | WMEMory<N>}. <N>, for chan-
nels, functions, TDR responses and waveform memories, is 1, 2, 3, or 4.
Mode
Oscilloscope and TDR modes. Eye mode uses this for average optical power measurements.
Query
:MEASure:SOURce?
The query returns the current source selection.
Returned Format
[:MEASure:SOURce] <source>[,<source>]<NL>
Example
10 OUTPUT 707;":MEASURE:SOURCE CHANNEL1"
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......