2-4
Sample Programs
C Programming Examples
void initialize ( )
{
write_IO ("*RST");
/* reset scope - initialize to known state */
write_IO ("*CLS");
/* clear status registers and output queue */
write_IO (":SYSTem:HEADer OFF"); /* turn off system headers */
/* initialize time base parameters to center reference, 2 ms full-scale (200 us/div), and 20 us delay */
write_IO (":TIMebase:REFerence CENTer;RANGe 2e-3;POSition 20e-6");
/* initialize Channel1 1.6V full-scale (200 mv/div); offset -400mv */
write_IO (":CHANnel1:RANGe 1.6;OFFSet -400e-3");
/* initialize trigger info: channel1 signal on positive slope at 300mv */
write_IO (":TRIGger:SOURce FPANel;SLOPe POSitive");
write_IO (":TRIGger:LEVel-0.40");
/* initialize acquisition subsystem */
/* Real time acquisition - no averaging; record length 4096 */
write_IO (":ACQuire:AVERage OFF;POINts 4096");
} /* end initialize ( ) */
/*
* Function name: acquire_data
* Parameters: none
* Return value: none
* Description: This routine acquires data according to the current instrument settings.
*/
void acquire_data ( )
{
/*
* The root level :DIGitize command is recommended for acquisition of new
* data. It will initialize data buffers, acquire new data, and ensure that
* acquisition criteria are met before acquisition of data is stopped.
* The captured data is then available for measurements, storage, or transfer
* to a PC. Note that the display is automatically turned off by the
* :DIGitize command and must be turned on to view the captured data.
*/
write_IO (":DIGitize CHANnel1");
write_IO (":CHANnel1:DISPlay ON");
/* turn on channel 1 display which is turned off by the :DIGitize command */
} /* end acquire_data() */
/*
* Function name: auto_measurements
* Parameters: none
Summary of Contents for 86100A
Page 18: ...1 14 Introduction Status Reporting Figure 1 4 Status Reporting Data Structures...
Page 19: ...1 15 Introduction Status Reporting Status Reporting Data Structures continued...
Page 94: ...2 40 Sample Programs BASIC Programming Examples...
Page 124: ...4 16 Root Level Commands VIEW...
Page 132: ...5 8 System Commands TIME...
Page 140: ...6 8 Acquire Commands SWAVeform RESet...
Page 152: ...7 12 Calibration Commands STATus...
Page 192: ...11 12 Display Commands SSAVer...
Page 202: ...12 10 Function Commands VERTical RANGe...
Page 203: ...13 AREA 13 2 DPRinter 13 2 FACTors 13 3 IMAGe 13 3 PRINters 13 4 Hardcopy Commands...
Page 316: ...21 12 TDR TDT Commands Rev A 05 00 and Below STIMulus...
Page 332: ......