Remote Commands
Setting the State of Polarization
70
Agilent
8169A
Polarization Controller, Second Edition
Example
OUTPUT 724;":CIRC:EPS 128"
OUTPUT 724;":CIRC:THET 270"
OUTPUT 724;":CIRC:EPS?"
ENTER 724;E$
OUTPUT 724;":CIRC:THET?"
ENTER 724;T$
[:INPut]:POSition:HALF
Syntax
[:INPut]:POSition:HALF <wsp> <value>MINimumMAXimumDEFault
where value is a oating point number between -360.00 and 360.00.
Description
This command sets the position of the
λ
/2 retarder plate. The parameter
may be either
• a number, in mechanical degrees (do not give a unit; the number will be rounded
to the nearest 0.05°),
• MINimum (-360.00°),
• MAXimum (360.00°), or
• DEFault (0.00°).
[:INPut]:POSition:HALF?
Syntax
[:INPut]:POSition:HALF?
Description
This query gets the position of the
λ
/2 retarder plate in mechanical
degrees (without a unit).