Insertion Loss Variation with Rotation of
λ
/4 and
λ
/2 Plates
Performance Test
Agilent
8169A
Polarization Controller, Second Edition
121
7
Get measurement results, MIN/MAX readings:
a
Press
to get SHOW.
b
Press
to get MAXIMUM.
c
Note the displayed value in the test record.
d
Press
to get MINIMUM.
e
Note the displayed value in the test record.
f
Press
to get SHOW.
8
Press
until you get LOGGING.
9
Repeat list item 6 to list item 7 for wavelengths 1510nm, 1530nm and 1560nm.
To change the wavelength on the 8153A:
a Press , ,
until
λ
appears.
b Use the Modify keys to set the appropriate
λ
value.
To change the wavelength on the 8168A:
c
Press
, press
, type the wavelength on the numeric
keypad, press
.
Always note the MAXIMUM and MINIMUM values in your test record.
10
Press
to get back to MEASURE Mode.
11
Calculate the difference between the \Maximum Power" and the “Minimum
Power" as the result for \Insertion Loss Variation with rotation of
λ
/4 and
λ
/2
plates".