Performance Test
Extinction Ratio of Polarizer
128
Agilent
8169A
Polarization Controller, Second Edition
3
Set up the 8153A:
a
Zero the 8153A: press
b
Set 8153A to Auto ranging: press
c
Set wavelength to 1470nm: press
until _ appears use Modify keys
to set the appropriate
λ
value.
d
Set display to dB: press
e
Set to averaging time 50ms: press
until T appears use Modify keys
to set the appropriate T value.
4
Set up 8168C:
a
Set wavelength to 1470nm : press
, type “1470", press
.
b Set power to 400_W: press
, type
“
400", press
.
c
Activate Optical Output.
5
Set up the 8169A:
a
Set to Home position: press
6
Adjust the tunable filter to get maximum transmission: 8153A display shall show
maximum reading (minimum value at negative sign).
7
Set 8169A (DUT) to maximum transmission:
a
Set the polarizing _lter for maximum transmission (see “Setting the Position
of the Polarizing Filter" in Chapter 2).
b
Set 8169A (DUT) to Circle: press
c Select 2
θ
P
and turn the knob to get maximum reading on 8153A display.
d Select 2
ε
Β
and turn knob to get maximum reading on 8153A display.
8
On the 8153A: press
.
9
Add 180 degrees to displayed value of 2
ε
Β
and enter this value: type new value,
press
10
Set 8169A (DUT) to minimum transmission:
a
Select 2
θ
P
and turn the knob to get minimum reading on 8153A display
(maximum value at negative sign).
b Select 2
ε
Β
and turn the knob to get minimum reading on 8153A display.
You should repeat this step several times using the smallest step size
when turning the knob.