Insertion Loss versus Wavelength
Performance Test
Agilent
8169A
Polarization Controller, Second Edition
125
12
Set the 8153A to wavelength 1470nm.
13
Note the displayed power on 8153A in your test records in the column
“Reference Power" for the associated wavelength.
14
Repeat list item 11 to list item 13 for the wavelengths from 1480 to 1570nm in
steps of 10nm, always setting the 8168A and the 8153A to the required
wavelength.
15
Repeat list item 11 to list item 13 again for wavelength 1470 nm to ensure
stability of measurement setup. If your measured value is more than .01 dB off
the previous value you need to fix your setup and repeat list item 11 to list
item 15.
16
Calculate “Insertion Loss" as difference of “Reference Power"- “Power after
DUT" and note the values in the associated column for each wavelength.
17
Check your calculations of “Insertion Loss" for the maximum and minimum
value and note the values on the associated lines. The maximum value applies
for Maximum Insertion Loss specification.
18
Calculate the difference of maximum and minimum value as the result of
“Variation of Insertion Loss with Wavelength".