GPIB Interface
Programming the Polarization Controller
Agilent
8169A
Polarization Controller, Second Edition
43
Setting the GPIB Address
You can only set the GPIB address from the front panel. See “Setting the
GPIB Address" in Chapter 4.
The default GPIB address is 24.
Returning the Instrument to Local Control
If the instrument has been operated in remote the only key you can use is
. The
key returns the instrument to local control.
does not operate if local lockout has been enabled.
Table 1 GPIB Capabilities
Mnemonic
Function
SH1
Complete source handshake capability
AH1
Complete acceptor handshake capability
T6
Basic talker; serial poll; no talk only mode; unaddressed to talk
if addressed to listen
L4
Basic listener; no listen only mode; unaddressed to listen if ad
dressed to talk
SR1
Complete service request capability
RL1
Complete remote/local capability
PP0
No parallel poll capability
DC1
Complete device clear capability
DT0
No device trigger capability
C0
No controller capability.