Programming the Polarization Controller
GPIB Interface
46
Agilent
8169A
Polarization Controller, Second Edition
Short Form and Long Form
The instrument accepts messages in short or long forms. For example, the
message :DISPLAY:ENABLE ON is in long form, the short form of this
message is DISP:ENAB ON.
In this manual the messages are written in a combination of upper and
lower case. Upper case characters are used for the short form of the
message. For example, the above command would be written
:DISPlay:ENABle.
The first colon can be left out for the first command or query in your
message. That is, the example given above could also be sent as
DISP:ENAB ON.